Automated Yumizen G1500 coagulation analyzer with 8 measurement channels and the ability to study lipemic samples
- In stock
- text_sku 9368
Automated Yumizen G1500 coagulation analyzer with 8 measurement channels and the ability to study lipemic samples meets those daily expectations thanks to its unique features
- Clotting, immunologic and chromogenic assays
- Integrated nephelometric and immuno-turbidimetric measuring channels
- Up to 8 measuring channels with 2 wavenlenght combination
- Continuous loading (sample racks, reagents, cuvettes)
- Real-time remote access
- Comprehensive system (reagents, QC, consumables, services support)
- Safety single use cuvette
- Touch screen
Price on request
System auto-verification
- 44 cooled controlled positions for reagents
- Up to 4 buffer and 8 control positions
- Liquid level sensor
- Positive barcode identification
- Up to 1000 cuvettes for testing and 4.5L of cleaner
Run Calibration / QC
- Pre calibrated tests
- Auto-control setup, customizable according to activity (number of tests, new reagents...)
- Levey-Jennings quality control char
Load Samples
- 150 samples / 15 racks
- Cap piercing (optional) for the Closed-tube mode
- 8 STAT sample positions
Run and Validation
- Real time curve monitoring
- Automated rerun and reflex testing
Dimensions & Weight
Length: 1000 mm, Height: 600 mm, Width: 900 mm, 85 kg
Inspection principle
Colorimetry - 405 nm; turbidimetry - 570 nm; nephelometry - 640 nm
Sample type
Plasma (3.2% sodium citrate)
Measurement channels
8 independent measurement channels
Sampling volume
min - 5 µl, max - 200 µl
Incubation
Built-in incubator 37.4 °C +/- 0.4 °C
Test speed
Up to 250 samples/day
Throughput
225 Prothrombin Time tests/hour
USB
6 x USB plugs
Calibration
Calibration curve (max. 6 points)
Special features
Warning system for possible measurement errors
System for reagents
44 cooled reagent positions
Quality Control (QC)
Levy-Jennings charts
Storage capacity
1,000,000 measurement data and corresponding curves (ID number, date and time of measurement)
Reaction cuvettes
Disposable cuvettes are loaded in bulk. Safe disposable cuvette for testing
System for samples
15 tripods with 10 positions for samples, 8 positions for STAT
Parameters
Clotting tests (640 nm);
Screening tests (PT, APTT, Fibrinogen, TF);
Factors (II, V, VII, X, VIII, IX, XI, XII);
D-Dimer turbidimetric test (570 nm);
Chromogenic test (405 nm) ATIII
Maximum measurement time
120 seconds
Введення даних
Liquid crystal color touchscreen display 270x340 mm. Integrated barcode reader for samples and reagents